Data Logger Manual
Page 3
Section 1: Device Safety
CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSIS-
TENT WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, MADGETECH SPECIFIES
USE OF AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY
ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM MADGETECH.
1.1 Model information
• Model:pHTemp2000
• Description:Measure&recordpHandtemperatureusingan
external probe with LCD display
• ManufacturedbyMadgeTech,Inc.Warner,NHUSA
1.2 Specifications
Temperature*
Measurement Range: -40 to 110C (80 to 144 ohms)
Resolution: 0.01C (0.001 ohms)
Calibrated Accuracy: ±0.1°C @ 25°C Ambient (±0.015 ohms)
Input Connection: RTD; Removable screw terminal; 2, 3, or 4-wire RTD
Probe**
pH
pH Range: 0~14pH (±1000mV)
pH Resolution: 0.01pH
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment
*Temperature specifications based on an ideal 100 ohm Pt RTD compliant with IEC 751 (1983)
and ITS-90, 5000 ohms FSR
**A compatible RTD probe is required to accurately measure pH